Navigation:SATA TECHNOLOGY CO., LTD >>Product >> Safety Instruments >> ST-2811 LCR Digital Bridge
Product

ST-2811 LCR Digital Bridge

Introduction

ST-2811 LCR Digital Bridge

Product Details

 

ST-2811 LCR DIGITAL BRIDGE
Technical parameters:
1.         Test frequency: 100Hz, 1kHz, 10kHz±0.02%
2.         Power of test signal: 0.3Vrms±15%(non-load) @ 1kHz
3.         Test item: L/Q, C/D, R/Q, and display by LED
4.         Measuring range
Items
Frequency
Range
L
100Hz, 120Hz
1uH~9999H
1kHz
0.1μH~999.9H
10kHz
0.01μH~99.99H
C
100Hz, 120Hz
1pF~9999μF
1kHz
0.1pF~999.9μF
10kHz
0.01pF~99.99μF
R
 
0.001Ω~99.99MΩ
Q
 
0.001~9999
D
 
0.0001~9.999
5.         Measuring accuracy:
Items
Frequency
Accuracy
L
100Hz120Hz
±[1μH+0.25%(1+L/200H+2mH/L)](1+1/Q)
1kHz
±[0.1μH+0.25%(1+L/200H+0.2mH/L)](1+1/Q)
10kHz
±[0.01μH+0.5%(1+L/10H+0.04mH/L)](1+1/Q)
C
100Hz120Hz
±[1pF+0.25%(1+1000pF/Cx+Cx/1000μF)](1+Dx)
1kHz
±[0.1pF+0.25%(1+100pF/Cx+Cx/100μF)](1+Dx)
10kHz
±[0.01pF+0.5%(1+20pF/Cx+Cx/4μF)](1+Dx)
R
 
±[1mΩ+0.25%(1+R/2MΩ+2Ω/R)](1+Q)
Q
100Hz,1kHz
±[0.020+0.15(Qx+1/Qx)%]
10kHz
±[0.020+0.2(Qx+1/Qx)%]
D
100Hz,1kHz
±0.0010(1+Dx2)
10kHz
±0.0015(1+Dx2)
6.         Test speed: 4~6 times/s
7.         Working environment: 0-40, and humidity is ≤85%RH
8.         Power supply: 220V/AC±10%, 50Hz±5%, power less than 30W
9.         Measurement: 360 * 125 * 320mm
10. Weight: Approx. 5kg

  • ©Copyright@SATA TECHNOLOGY CO., LTD
  • Tel:+86 769 83814636 FAX:+86 769 83509758
  • Address:Changping Town, Dongguan City, Guangdong Province, China 523560
Keywords:LCR Digital Bridge, LCR meter, LCR Tester